I to be unsure about the accuracy and/or precision of measure up instruments. For example, if we have a i graduated cylinder, a pipette and also a beaker, how and why can we assert them the they deserve to be supplied with accuracy, precision, or both?


*

*

Beakers, Erlenmeyer flasks, dropping funnels and the like are not volumetric instruments. They space not exactly calibrated, and their scale serve just as almost right guides. Therefore, the beaker that is discussed in the concern is appropriate out.

You are watching: Accuracy of burette pipette and measuring cylinder

Graduated cylinders, volumetric flasks, bulb pipettes, i graduated pipettes, and burettes space volumetric instruments. They are calibrated; however, like any kind of other measure up instrument, lock only have a minimal accuracy and precision.

The accuracy and also precision carry out not only depend ~ above the type of the instrument (e.g. Graduated cylinder or pipette). They likewise depend on the dimensions, graduating divisions, and accuracy class of the instrument, which room usually around the world standardized.

In principle, volumetric tools with a small diameter have a far better precision because the same difference of the fluid level (i.e. The meniscus) synchronizes to a smaller distinction in volume. Therefore, usual long and also thin pipettes space usually better than short and also thick i graduated cylinders. Because that example:

A standardized i graduated pipette (Class A or AS) through a nominal volume of 10 ml and an as whole length of around 360 mm has a best permissible error the ±0.05 ml.

A standardized graduated measuring cylinder (Class A) through a nominal capacity of 10 ml and also a maximum all at once height that 140 mm (tall form) has actually a preferably permissible error the ±0.1 ml.

If you use an also larger i graduated measuring cylinder (Class A) through a nominal capacity of 100 ml and fill it simply to the 10 ml mark, the best permissible error is ±0.5 ml.

These errors stand for the best permissible error at any allude on the scale, and additionally the maximum permissible difference between the errors at any type of two points.

Since the very same sources the error are, naturally, natural both in calibration and use, a volumetric tool should have actually a good precision in bespeak to be able to be calibrated v a great accuracy.

See more: What Does An Acid Base Titration Accomplish, Lab Procedure: Titration

However, keep in mind that i graduated pipettes room usually calibrated ‘to deliver’ (TD, Ex), whereas i graduated measuring cylinders space calibrated ‘to contain’ (TC, In). Hence, graduatted pipettes and graduated measure cylinders might not be interchangeable interchangeable at will.